GL260 Main unit specifications
| Item | Description |
|---|---|
| Number of analog input channels | 10 channels |
| External input/output – Input | Trigger or Sampling (1 channel), Logic/Pulse (4 channels) Voltage range: Up to 24 V (common ground) Signal type: Voltage, Open collector, Contact (relay) Threshold: Approx. +2.5 V (Hysteresis: Approx. 0.5 V [2.5 V to 3 V]) |
| External input/output – Output | Alarm (4 channels) Output signal: Open collector (pull-up to 5 V by 10 kΩ resistor) *Maximum rating of the output transistor* Voltage: Max. 30 V Current: Max. 0.5 A Collector dissipation: Max. 0.2 W |
| Sampling interval | 10 ms to 1 hour (10 ms to 50 ms: voltage only. Minimum interval varies by number of channels used.) External signal input |
| Time scale of waveform display | 1 sec. to 24 hour/division |
| Trigger – Repeat action | Off, On (auto rearmed) |
| Trigger – Trigger action | Start or stop capturing data by the trigger |
| Trigger – Trigger source | Start: Off, Measured signal, Alarm, External, Clock, Week or Time Stop: Off, Measured signal, Alarm, External, Clock, Week or Time |
| Alarm function – Condition setting | Combination: OR or AND Analog signal: Rising (High), Falling (Low), Window-in, Window-out Logic signal: Pattern (combination of each input signal in high or low) Pulse (number of count): Rising (High), Falling (Low), Window-in, Window-out |
| Pulse input – Measurement mode | Rotation count (RPM) mode |
| Pulse input – Maximum number of pulse inputs | Maximum input frequency: 5 kHz Maximum number of count: 50k/sampling (16-bit counter) |
| Interface to PC | USB (Hi-speed), WLAN (using B-568 option) |
| Storage device – Internal | Built-in 8 GB Flash Memory |
| Storage device – External | One SD card slot (Support SDHC memory card, up to 32 GB) |
| Storage device – Saved contents | Captured data, Setting conditions, Screen copy |
| Display | 4.3-inch TFT color LCD (WQVGA: 480 × 272 dots) |
| Operating environment | 0 to 45 °C, 5 to 85 % RH (non condensed) (When operating with battery pack: 0 to 40 °C, charging battery 15 to 35 °C) |
| Power source – AC adapter | AC100–240 V / 50–60 Hz |
| Power source – DC power | DC8.5–24 V |
| Power source – Battery pack | Mountable battery pack (battery pack option B-573: 7.2 V DC, 2875 mAh) |
| Power consumption | Max. 36 VA (Rating under maximum power consumption using the AC adapter, with LCD display on, and battery pack being charged.) |
| External dimensions (W × D × H) | Approx. 188 × 117 × 42 mm (excluding projections) |
| Weight | Approx. 500 g (excludes AC adapter and battery pack) |
| Vibration-resistant conditions | Equivalent to equipment of automobile parts Type 1 Class A |
GL260 Analog input specifications
| Item | Description |
|---|---|
| Type of input terminal | M3 screw type terminals (rectangular flat washer) |
| Input method | Photo MOS relay scanning system, All channels isolated, balanced input |
| Measurement range – Voltage | 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 100 V; 1–5 V F.S. |
| Measurement range – Thermocouple | K, J, E, T, R, S, B, N, C (W: WRe5-26) |
| Measurement range – Humidity | 0 to 100% (voltage 0 to 1 V scaling conversion) – using the humidity sensor (option B-530) |
| Filter | Off, 2, 5, 10, 20, 40 (moving average in selected number) |
| A/D converter | Sigma-Delta type, 16 bits (effective resolution: 1/40000 of the measuring full range) |
| Maximum input voltage – Between +/- terminals | 20 mV to 1 V range: 60 Vp-p 2 V to 100 V range: 110 Vp-p |
| Maximum input voltage – Between input terminal / input terminal | 60 Vp-p |
| Maximum input voltage – Between input terminal / GND | 60 Vp-p |
Software specifications for PC
| Item | Description |
|---|---|
| Model name | GL28-APS |
| Supported OS | Windows 11 (64bit) |
| Functions | Control the GL series, Real-time data capture, and Data format conversion |
| Number of groups | Up to 4 groups |
| Number of CHs per 1 group | Up to number of connected module |
| Maximum number of channels | 1000 ch maximum |
| Settings control | AMP settings, capture settings, trigger/alarm settings, report settings, others |
| Capturing data | Real-time data (CSV, GBD binary) Data in internal memory or SD memory card (CSV, GBD binary) |
| Display | Analog waveforms, logic waveforms, pulse waveforms, digital values |
| Display modes | Y-T View, Digital View, X-Y View between Cursors (only during replay) |
| File conversion | Between cursors, All data |
| Monitor functions | Alarm monitor enables sending of email to the specified address |
| Statistic/History | Displays maximum, minimum and average values during measurement |
| Report Function | Enables creation of daily or monthly files |
Standard accessories
| Item | Description |
|---|---|
| Notice | This document provides a clear explanation of the basic operations |
| TO ENSURE SAFE AND CORRECT USE | This document contains important information to ensure the safe and proper use of the product |
| AC adapter | 100 to 240 VAC, 50/60 Hz, Power supply cord for each area |
| Ferrite core | Attach to cable for radiation reduction |
Wireless LAN unit (option) specifications
| Item | Description |
|---|---|
| Model number | B-568 |
| Communication method | Wireless communication (using radio waves in the 2.4 GHz band) |
| Installation | Attach to the wireless LAN connection terminal |
| Wireless LAN standard | IEEE802.11b/g/n |
| Function | Control from PC, data transfer to PC, control and data transfer from smartphone/tablet - For access point: Local control and transfer are possible. - For station: In addition to local, remote control and transfer using a REMOTE site is also possible. Communication range: Approx. 40 m Encryption function: WEP64, WEP128, WPA-PSK/WPA2-PSK (TKIP/AES) |
Humidity Sensor B-530 (option) Specifications
| Item | Description |
|---|---|
| Allowable temperature range | -25 to 80 °C |
| Allowable humidity range | 0 to 100% RH |
| Relative humidity measurement accuracy | ±3% RH (5 to 98% RH at 25 °C) |
| Method | Capacitance method |
| Relative humidity measurement accuracy (5 to 98%) | 0–10 °C: ±5% RH 10–20 °C: ±4% RH 20–30 °C: ±3% RH 30–40 °C: ±4% RH 40–50 °C: ±5% RH 50–60 °C: ±6% RH 60–70 °C: ±7% RH 70–80 °C: ±8% RH * Measurement accuracy at 60 °C or more is a reference value. |
| Response time | 15 sec. (90% response when membrane filter is installed) |
| Sensor output | DC0–1 V |
| External dimensions | Φ14 × 80 mm (excluding cable) |
| Cable length | 3 m |
| Sensor power source | DC5–16 V ±5% |
| Power consumption | Approx. 4 mA |
-
GL260 데이터 시트
- 확장자 :
- 파일크기 :
- 1.83MB